Metraipolis › archive › dc_ffb88eb40ec6

Evaluations of a Detector-Limited Digital Impedance Bridge — Europe PMC PMID:36475083

Source
Europe PMC (en)
Original
https://europepmc.org/article/MED/36475083
DOI
10.6028/jres.126.006
License (as archived)
CC0-1.0
Archived
2026-11-05T03:36:17.454Z
SHA-256 of the archived text
28857a7b0e2c50d0d003b14c212402977c667c78f7cacac625720d3d9fc9efaf
Stable URL of this page
https://metraipolis.com/documento/dc_ffb88eb40ec6

How to cite

Evaluations of a Detector-Limited Digital Impedance Bridge — Europe PMC PMID:36475083 — Europe PMC (en), https://europepmc.org/article/MED/36475083 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_ffb88eb40ec6 (Metraipolis archive dc_ffb88eb40ec6, sha256 28857a7b0e2c50d0d003b14c212402977c667c78f7cacac625720d3d9fc9efaf)
BibTeX
@misc{dcffb88eb40ec6,
  title = {Evaluations of a Detector-Limited Digital Impedance Bridge — Europe PMC PMID:36475083},
  howpublished = {Europe PMC (en)},
  url = {https://europepmc.org/article/MED/36475083},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_ffb88eb40ec6; sha256 28857a7b0e2c50d0d003b14c212402977c667c78f7cacac625720d3d9fc9efaf},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/36475083. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

We tested a simple digital impedance bridge using two nominally equal resistors to form a 1:1 ratio. We focused on resolution and stability of the detectors. Fluctuations of the source voltages were largely removed through postprocessing of the digitized data, and the measurement results were limited by the detector noise. This detector-limited operating condition was first demonstrated using three modified Keysight 3458A multimeters for measurements of the voltage ratios, achieving 0.01 μV/V type A uncertainty in less than 15 min at 1 kHz. In an effort to extend the applicable frequency range and develop a system with off-the-shelf components, we tested a system using three lock-in detectors for measuring small deviations from the perfect AC ratio of unity magnitude, achieving stabilities and resolutions of 0.1 μV/V in a few hours for each point from 1 kHz to 5 kHz.

Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera).
Pubblicazione: Journal of research of the National Institute of Standards and Technology, 2021 (prima pubblicazione 2021-04-06).
Identificativi: PMID 36475083 · PMCID PMC9681207 · DOI 10.6028/jres.126.006.
Parole chiave degli autori: Noise Cancellation; Impedance Standard; Ac Voltage Ratio; Digital Bridge; Lock-In Detector.
Pagina: https://europepmc.org/article/MED/36475083 — https://doi.org/10.6028/jres.126.006. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0.
Letto il 2026-11-05T03:31:22.801Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.