Metraipolis › archive › dc_fd50a6338633

Drape-Compatible Tool-Tip Localization for Hand-Held Laparoscopic Instruments via UWB Carrier-Phase Ranging and Trocar-Constrained Geometry — arXiv:2609.39129

Source
arXiv (en)
Original
https://arxiv.org/abs/2609.39129
License (as archived)
CC0-1.0
Archived
2026-10-07T15:11:55.052Z
SHA-256 of the archived text
9641e391e47241de886859b1b5cd0c8b47d43e891c05d9643ce81d7b68832145
Stable URL of this page
https://metraipolis.com/documento/dc_fd50a6338633

How to cite

Drape-Compatible Tool-Tip Localization for Hand-Held Laparoscopic Instruments via UWB Carrier-Phase Ranging and Trocar-Constrained Geometry — arXiv:2609.39129 — arXiv (en), https://arxiv.org/abs/2609.39129 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_fd50a6338633 (Metraipolis archive dc_fd50a6338633, sha256 9641e391e47241de886859b1b5cd0c8b47d43e891c05d9643ce81d7b68832145)
BibTeX
@misc{dcfd50a6338633,
  title = {Drape-Compatible Tool-Tip Localization for Hand-Held Laparoscopic Instruments via UWB Carrier-Phase Ranging and Trocar-Constrained Geometry — arXiv:2609.39129},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2609.39129},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_fd50a6338633; sha256 9641e391e47241de886859b1b5cd0c8b47d43e891c05d9643ce81d7b68832145},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2609.39129. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

A surgical robot policy needs to know where each instrument's working end sits relative to the camera and to the other instrument, yet a laparoscopic operation leaves only the endoscope video, and a draped instrument hides every optical path on itself. We estimate the tool tips from distances and an IMU alone. The distances are measured by ultra-wideband carrier phase between antenna nodes on the handle side of the instruments: one per hand-held instrument, one on the endoscope, all outside the sterile barrier. Take the endoscope antenna as the reference. The two instrument antennas carry six coordinates while the three pairs give three distances, so the problem is short by three, and averaging cannot close that gap because what is missing is information, not precision. Carrier phase adds one more unknown per pair, an integer that leaves distance fixed only modulo lambda/2 = 23.1 mm. The trocar settles both difficulties: each shaft passes through a port whose position is known and whose direction the on-board IMU measures, so its antenna keeps a single degree of freedom, the insertion depth. Two unknowns now face three distances -- two fix the depths and the third is left over, and that spare equation exposes a wrong integer or a wrong mounting offset instead of absorbing it. The same solve returns both tool tips in the endoscope frame, without the second carrier frequency that integer resolution usually needs, so the link never leaves a single PLL lock. In an 11-hole phantom among metal instruments the three distances hold sigma = 0.10-0.33 mm at 35-46 Hz, a sterile drape pressed onto the antennas leaves the link unchanged where an optical path would be blocked, and the logger was carried into a live rabbit appendectomy.

Tema degli esploratori: statistica e scienza dei dati (retro-catalogo arXiv: annunci del 2026-10-01 in cs.RO).
Categorie arXiv: Robotics (cs.RO).
Pagina del lavoro: https://arxiv.org/abs/2609.39129 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina.
Letto il 2026-10-07T14:44:47.418Z dalla pagina https://arxiv.org/catchup/cs.RO/2026-10-01?abs=True (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).