Metraipolis › archive › dc_f6f7de185bca
Sensor orientation and other factors which increase the blast overpressure reporting errors — Europe PMC PMID:33031423
- Source
- Europe PMC (en)
- Original
- https://europepmc.org/article/MED/33031423
- DOI
- 10.1371/journal.pone.0240262
- License (as archived)
- CC0-1.0
- Archived
- 2026-11-07T08:48:25.117Z
- SHA-256 of the archived text
8dc99f2ede1dac3f0b3f1c0e07ba457b26e5142564172fa436c5d907ee39505f- Stable URL of this page
https://metraipolis.com/documento/dc_f6f7de185bca
How to cite
Sensor orientation and other factors which increase the blast overpressure reporting errors — Europe PMC PMID:33031423 — Europe PMC (en), https://europepmc.org/article/MED/33031423 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_f6f7de185bca (Metraipolis archive dc_f6f7de185bca, sha256 8dc99f2ede1dac3f0b3f1c0e07ba457b26e5142564172fa436c5d907ee39505f)
BibTeX
@misc{dcf6f7de185bca,
title = {Sensor orientation and other factors which increase the blast overpressure reporting errors — Europe PMC PMID:33031423},
howpublished = {Europe PMC (en)},
url = {https://europepmc.org/article/MED/33031423},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_f6f7de185bca; sha256 8dc99f2ede1dac3f0b3f1c0e07ba457b26e5142564172fa436c5d907ee39505f},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/33031423. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
This study compared the response of the wearable sensors tested against the industry-standard pressure transducers at blast overpressure (BOP) levels typically experienced in training. We systematically evaluated the effects of the sensor orientation with respect to the direction of the incident shock wave and demonstrated how the averaging methods affect the reported pressure values. The evaluated methods included averaging peak overpressure and impulse of all four sensors mounted on a helmet, taking the average of the three sensors, or isolating the incident pressure equivalent using two sensors. The experimental procedures were conducted in controlled laboratory conditions using the shock tube, and some of the findings were verified in field conditions with live fire charges during explosive breaching training. We used four different orientations (0°, 90°, 180°, and 270°) of the headform retrofitted with commonly fielded helmets (ACH, ECH, Ops-Core) with four B3 Blast Gauge sensors. We determined that averaging the peak overpressure values overestimates the actual dosage experienced by operators, which is caused by the reflected pressure contribution. This conclusion is valid despite the identified limitation of the B3 gauges that consistently underreport the peak reflected overpressure, compared to the industry-standard sensors. We also noted consistent overestimation of the impulse. These findings demonstrate that extreme caution should be exercised when interpreting occupational blast exposure results without knowing the orientation of the sensors. Pure numerical values without the geometrical, training-regime specific information such as the position of the sensors, the distance and orientation of the trainee to the source of the blast wave, and weapon system used will inevitably lead to erroneous estimation of the individual and cumulative blast overpressure (BOP) dosages. Considering that the 4 psi (~28 kPa) incident BOP is currently accepted as the threshold exposure safety value, a misinterpretation of exposure level may lead to an inaccurate estimation of BOP at the minimum standoff distance (MSD), or exclusion criteria. Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera). Pubblicazione: PloS one, 2020 (prima pubblicazione 2020-10-08). Identificativi: PMID 33031423 · PMCID PMC7544144 · DOI 10.1371/journal.pone.0240262. Descrittori MeSH principali: Explosions; Pressure; Research Design; Wearable Electronic Devices. Pagina: https://europepmc.org/article/MED/33031423 — https://doi.org/10.1371/journal.pone.0240262. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0. Letto il 2026-11-07T08:43:51.850Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.