Metraipolis › archive › dc_d9f23807fa27

OpenULF: A Modular Toolkit Enables Data-Driven Ultra-Low-Field MRI with both Classical and Quantum Sensors — arXiv:2611.01511

Source
arXiv (en)
Original
https://arxiv.org/abs/2611.01511
License (as archived)
CC0-1.0
Archived
2026-11-03T18:56:50.406Z
SHA-256 of the archived text
1816ab8290a514604b2ee223c556b0f4d2f709e2efad2a114f0600750b1d3f5a
Stable URL of this page
https://metraipolis.com/documento/dc_d9f23807fa27

How to cite

OpenULF: A Modular Toolkit Enables Data-Driven Ultra-Low-Field MRI with both Classical and Quantum Sensors — arXiv:2611.01511 — arXiv (en), https://arxiv.org/abs/2611.01511 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_d9f23807fa27 (Metraipolis archive dc_d9f23807fa27, sha256 1816ab8290a514604b2ee223c556b0f4d2f709e2efad2a114f0600750b1d3f5a)
BibTeX
@misc{dcd9f23807fa27,
  title = {OpenULF: A Modular Toolkit Enables Data-Driven Ultra-Low-Field MRI with both Classical and Quantum Sensors — arXiv:2611.01511},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2611.01511},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_d9f23807fa27; sha256 1816ab8290a514604b2ee223c556b0f4d2f709e2efad2a114f0600750b1d3f5a},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2611.01511. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Ultra-low-field (ULF) Magnetic Resonance Imaging (MRI) offers a portable, low-cost alternative to conventional scanners by operating at field strengths below 10 mT. At these fields, two sensor technologies compete: quantum SQUID (Superconducting Quantum Interference Device) sensors, which provide high sensitivity but require cryogenic cooling, and classical induction coils, which are inexpensive and room-temperature but produce weaker signals. With conventional physics-based imaging, both suffer from low signal-to-noise ratios (SNR). Data-driven reconstruction has overcome similar limits in related inverse problems, but its use for ULF-MRI has been hindered by a lack of open datasets, benchmarks, and flexible simulators. To close this gap, we present OpenULF, a modular, open-source simulation toolkit that generates paired training data and supports end-to-end machine-learning (ML) reconstruction. OpenULF integrates a Bloch-equation-validated physics simulator, calibrated SQUID and coil receiver models, reconstruction ML networks, and an automated design-space exploration engine. Architecturally, a receiver abstract base class, a typed HDF5 contract that decouples physics from learning, and a resumable sweep engine make sensors, models, and datasets independently substitutable. Using OpenULF, we train 774 models across four anatomical datasets and conduct three design-space explorations. The effectiveness of ML reconstruction depends strongly on the dataset and the sensor. We identify a SQUID-to-coil crossover near 450 $\mu$T and demonstrate that sensor hardware and ML model capacity can be traded off during system design. Sensor and reconstruction choices must therefore be co-designed: the best sensor is set by the dataset and field strength, not by SNR.

Tema degli esploratori: statistica e scienza dei dati (elenco arXiv cs.CV, nuovi lavori del giorno).
Categorie arXiv: Image and Video Processing (eess.IV) ; Computer Vision and Pattern Recognition (cs.CV); Medical Physics (physics.med-ph).
Pagina del lavoro: https://arxiv.org/abs/2611.01511 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina.
Letto il 2026-11-03T18:16:53.108Z dalla pagina https://arxiv.org/list/cs.CV/new (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).