Metraipolis › archive › dc_d3f6fd4c07a1
Optical Power Scale Realization by Laser Calorimeter after 45 Years of Operation — Europe PMC PMID:38469451
- Source
- Europe PMC (en)
- Original
- https://europepmc.org/article/MED/38469451
- DOI
- 10.6028/jres.126.011
- License (as archived)
- CC0-1.0
- Archived
- 2026-11-03T21:46:11.780Z
- SHA-256 of the archived text
64cfd16b21de7dfcc1ae611d32f0f8a2380ec1c245491f8192924a5d103f36d4- Stable URL of this page
https://metraipolis.com/documento/dc_d3f6fd4c07a1
How to cite
Optical Power Scale Realization by Laser Calorimeter after 45 Years of Operation — Europe PMC PMID:38469451 — Europe PMC (en), https://europepmc.org/article/MED/38469451 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_d3f6fd4c07a1 (Metraipolis archive dc_d3f6fd4c07a1, sha256 64cfd16b21de7dfcc1ae611d32f0f8a2380ec1c245491f8192924a5d103f36d4)
BibTeX
@misc{dcd3f6fd4c07a1,
title = {Optical Power Scale Realization by Laser Calorimeter after 45 Years of Operation — Europe PMC PMID:38469451},
howpublished = {Europe PMC (en)},
url = {https://europepmc.org/article/MED/38469451},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_d3f6fd4c07a1; sha256 64cfd16b21de7dfcc1ae611d32f0f8a2380ec1c245491f8192924a5d103f36d4},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/38469451. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
To calibrate laser power and energy meters, the National Institute of Standards and Technology (NIST) uses several detector-based realizations of the scale for optical radiant flux; these realizations are appropriate for specific laser power/energy ranges and optical coupling configurations. Calibrations from 1 µW to 2 W are currently based upon calorimeters. Validation by comparisons against other primary representations of the optical watt over the last two decades suggests the instruments operate well within their typical reported uncertainty level of 0.86 % with 95 % confidence. The dominant uncertainty contribution in the instrument is attributable to light scattered by the legacy window, which was not previously recognized. The inherent electro-optical inequivalence in the calorimeter's response was reassessed by thermal modeling to be 0.03 %. The principal contributions to the overall inequivalence were corrected, yielding a shift in scale representation under 0.2 % for typical calibrations. With updates in several uncertainty contributions resulting from this reassessment, the resulting combined expanded uncertainty (k = 2) is 0.84 %, which is essentially unchanged from the previous result provided to calibration customers. Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera). Pubblicazione: Journal of research of the National Institute of Standards and Technology, 2021 (prima pubblicazione 2021-06-28). Identificativi: PMID 38469451 · PMCID PMC10046752 · DOI 10.6028/jres.126.011. Parole chiave degli autori: Calorimetry; Laser Power; Primary Standard. Pagina: https://europepmc.org/article/MED/38469451 — https://doi.org/10.6028/jres.126.011. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0. Letto il 2026-11-03T21:39:02.743Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.