Metraipolis › archive › dc_bc1353b94c7f
Source apportionment and quantification of liquid and headspace leaks from closed system drug-transfer devices via Selected Ion Flow Tube Mass Spectrometry… — Europe PMC PMID:34735484
- Source
- Europe PMC (en)
- Original
- https://europepmc.org/article/MED/34735484
- DOI
- 10.1371/journal.pone.0258425
- License (as archived)
- CC0-1.0
- Archived
- 2026-11-01T16:01:40.521Z
- SHA-256 of the archived text
271cd42122ce6200bbed685d10ec20ab1887f23895498e175bf93c3114c654f3- Stable URL of this page
https://metraipolis.com/documento/dc_bc1353b94c7f
How to cite
Source apportionment and quantification of liquid and headspace leaks from closed system drug-transfer devices via Selected Ion Flow Tube Mass Spectrometry… — Europe PMC PMID:34735484 — Europe PMC (en), https://europepmc.org/article/MED/34735484 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_bc1353b94c7f (Metraipolis archive dc_bc1353b94c7f, sha256 271cd42122ce6200bbed685d10ec20ab1887f23895498e175bf93c3114c654f3)
BibTeX
@misc{dcbc1353b94c7f,
title = {Source apportionment and quantification of liquid and headspace leaks from closed system drug-transfer devices via Selected Ion Flow Tube Mass Spectrometry… — Europe PMC PMID:34735484},
howpublished = {Europe PMC (en)},
url = {https://europepmc.org/article/MED/34735484},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_bc1353b94c7f; sha256 271cd42122ce6200bbed685d10ec20ab1887f23895498e175bf93c3114c654f3},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/34735484. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
A system to differentiate and quantify liquid and headspace vapor leaks from closed system drug-transfer devices (CSTDs) is presented. CSTDs are designed to reduce or eliminate hazardous drug (HD) exposure risk when compounding and administering HDs. CSTDs may leak liquid, headspace, or a mixture of the two. The amount of HD contained in liquid and headspace leaks may be substantially different. Use of a test solution containing two VOCs with differences in ratios of VOC concentrations in the headspace and liquid enables source apportionment of leaked material. SIFT-MS was used to detect VOCs from liquid and headspace leaks in the vapor phase. Included in this report is a novel method to determine the origin and magnitude of leaks from CSTDs. A limit of leak detection of 24 μL of headspace vapor and 0.14 μL of test liquid were found using Selected Ion Flow Tube Mass Spectrometry (SIFT-MS). Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera). Pubblicazione: PloS one, 2021 (prima pubblicazione 2021-11-04). Identificativi: PMID 34735484 · PMCID PMC8568112 · DOI 10.1371/journal.pone.0258425. Descrittori MeSH principali: Protective Devices. Pagina: https://europepmc.org/article/MED/34735484 — https://doi.org/10.1371/journal.pone.0258425. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0. Letto il 2026-11-01T15:53:50.710Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.