Metraipolis › archive › dc_ab4f99054eb0

Technical considerations for medical device manufacturers when designing gastrostomy tubes (G-tubes) using the new ISO 80369-3 connector — Europe PMC PMID:32730360

Source
Europe PMC (en)
Original
https://europepmc.org/article/MED/32730360
DOI
10.1371/journal.pone.0236644
License (as archived)
CC0-1.0
Archived
2026-11-07T21:57:08.517Z
SHA-256 of the archived text
b3548b10931c8e9cf5a69a90d0c3dcb7602e8c490f87e755037cacf7cdfe20fd
Stable URL of this page
https://metraipolis.com/documento/dc_ab4f99054eb0

How to cite

Technical considerations for medical device manufacturers when designing gastrostomy tubes (G-tubes) using the new ISO 80369-3 connector — Europe PMC PMID:32730360 — Europe PMC (en), https://europepmc.org/article/MED/32730360 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_ab4f99054eb0 (Metraipolis archive dc_ab4f99054eb0, sha256 b3548b10931c8e9cf5a69a90d0c3dcb7602e8c490f87e755037cacf7cdfe20fd)
BibTeX
@misc{dcab4f99054eb0,
  title = {Technical considerations for medical device manufacturers when designing gastrostomy tubes (G-tubes) using the new ISO 80369-3 connector — Europe PMC PMID:32730360},
  howpublished = {Europe PMC (en)},
  url = {https://europepmc.org/article/MED/32730360},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_ab4f99054eb0; sha256 b3548b10931c8e9cf5a69a90d0c3dcb7602e8c490f87e755037cacf7cdfe20fd},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/32730360. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Background: Gastrostomy tubes (G-tubes) are typically used when people cannot eat food by mouth. The connector section that allows G-tubes to connect to other devices, such as feeding sets or syringes, has been modified on some of the devices to reduce misconnections in hospital settings. The narrow internal diameter of the new connector, standardized under ISO 80369-3, has caused some users to express concern about a reduced flow rate. Previous studies performed on commercial devices determined that it was not conclusive how much the ISO 80369-3 connector contributed towards the reduced flow rate, because when manufacturers designed these new connector-based devices, they often changed other geometric variables (such as distal tube diameter, or length) at the same time. Thus, it became difficult isolating the effect of the connector from other geometric variables. Method: The key objective of this study was to investigate how different design variables impacted the flow rate through the G-tubes. 3D-printed devices were used to assess the geometric parameters in a systematic manner. Commercial diets and Newtonian analog fluids with matched viscosities were used for testing. Results: The flow path length of the "transition section" encompassing the standardized ISO 80369-3 connector in the new devices was found to cause reduced flow. Additionally, results showed that a shortened (≤ 10 mm) transition section, along with a 10% increase in the distal inner diameter of large bore devices (e.g., 24 Fr), can restore flow rates to levels consistent with the previous devices prior to the connector standardization. Conclusions: The strategy for restoring flow rates to previous levels may help alleviate concerns raised by multiple stakeholders such as health care professionals, patients, caregivers and device manufacturers. In addition, the approach proposed here can be used as a tool for designing future G-tube devices.

Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera).
Pubblicazione: PloS one, 2020 (prima pubblicazione 2020-07-30).
Identificativi: PMID 32730360 · PMCID PMC7392286 · DOI 10.1371/journal.pone.0236644.