Comparison of conventional versus customised Eurosil-4 Pink bolus for radiotherapy of the chest wall — Europe PMC PMID:35511918
- Source
- Europe PMC (en)
- Original
- https://europepmc.org/article/MED/35511918
- DOI
- 10.1371/journal.pone.0267741
- License (as archived)
- CC0-1.0
- Archived
- 2026-10-07T23:41:44.544Z
- SHA-256 of the archived text
76903c7d36168d603cc9416d79c6c3b1a300b33ee62e44bbe3e961df26cb8b00- Stable URL of this page
https://metraipolis.com/documento/dc_aaa4b625e39f
How to cite
Comparison of conventional versus customised Eurosil-4 Pink bolus for radiotherapy of the chest wall — Europe PMC PMID:35511918 — Europe PMC (en), https://europepmc.org/article/MED/35511918 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_aaa4b625e39f (Metraipolis archive dc_aaa4b625e39f, sha256 76903c7d36168d603cc9416d79c6c3b1a300b33ee62e44bbe3e961df26cb8b00)
BibTeX
@misc{dcaaa4b625e39f,
title = {Comparison of conventional versus customised Eurosil-4 Pink bolus for radiotherapy of the chest wall — Europe PMC PMID:35511918},
howpublished = {Europe PMC (en)},
url = {https://europepmc.org/article/MED/35511918},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_aaa4b625e39f; sha256 76903c7d36168d603cc9416d79c6c3b1a300b33ee62e44bbe3e961df26cb8b00},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/35511918. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
Introduction: In radiotherapy, the presence of air gaps near a tumour can lead to underdose to the tumour. In this study, the impact of air gaps on dose to the surface was evaluated. 3D-printing was used to construct a Eurosil-4 Pink bolus customised to the patient and its dosimetric properties were compared with that of Paraffin wax bolus. Methods: Surface dose was measured for flat sheets of Eurosil-4 Pink bolus with different thicknesses. Different air gap thicknesses were inserted between the bolus and the surface, and dose was measured for each air gap using 10 cm × 10 cm fields. This was repeated with the effective field size calculated from the patient plan. Surface dose was measured for varying angles of incidence. A customised chest phantom was used to compare dose for two customised Eurosil-4 Pink boluses, and commonly used Paraffin wax bolus. Results: The surface dose was found to be highest for 1.1 cm thick bolus. The decrease in surface dose for the Eurosil-4 Pink bolus was minimal for the 10 cm × 10 cm field, but higher for the effective field size and larger angles of incidence. For instance, the dose was reduced by 6.2% as a result of 1 cm air gap for the effective field size and 60 degree angle of incidence. The doses measured using Gafchromic film under the customised Eurosil-4 Pink boluses were similar to that of the Paraffin wax bolus, and higher than prescribed dose. Conclusions: The impact of air gaps can be significant for small field sizes and oblique beams. A customised Eurosil-4 Pink bolus has promising physical and dosimetric properties to ensure sufficient dose to the tumour, even for treatments where larger impact of air gaps is suspected. Tema degli esploratori: fisica e astronomia (ricerca Europe PMC «particle physics», solo open access). Pubblicazione: PloS one, 2022 (prima pubblicazione 2022-05-05). Identificativi: PMID 35511918 · PMCID PMC9071134 · DOI 10.1371/journal.pone.0267741. Descrittori MeSH principali: Thoracic Wall; Neoplasms. Pagina: https://europepmc.org/article/MED/35511918 — https://doi.org/10.1371/journal.pone.0267741. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0. Letto il 2026-10-07T23:22:56.886Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.