Metraipolis › archive › dc_97240c9341ff

Long-Term Stability of One-Inch Condenser Microphones Calibrated at the National Institute of Standards and Technology — Europe PMC PMID:26958445

Source
Europe PMC (en)
Original
https://europepmc.org/article/MED/26958445
DOI
10.6028/jres.120.012
License (as archived)
CC0-1.0
Archived
2026-10-09T22:41:30.256Z
SHA-256 of the archived text
ecd0cc5427210b336ea55c283d983aee70bf7c3be8797387d11a3da0edf8b610
Stable URL of this page
https://metraipolis.com/documento/dc_97240c9341ff

How to cite

Long-Term Stability of One-Inch Condenser Microphones Calibrated at the National Institute of Standards and Technology — Europe PMC PMID:26958445 — Europe PMC (en), https://europepmc.org/article/MED/26958445 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_97240c9341ff (Metraipolis archive dc_97240c9341ff, sha256 ecd0cc5427210b336ea55c283d983aee70bf7c3be8797387d11a3da0edf8b610)
BibTeX
@misc{dc97240c9341ff,
  title = {Long-Term Stability of One-Inch Condenser Microphones Calibrated at the National Institute of Standards and Technology — Europe PMC PMID:26958445},
  howpublished = {Europe PMC (en)},
  url = {https://europepmc.org/article/MED/26958445},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_97240c9341ff; sha256 ecd0cc5427210b336ea55c283d983aee70bf7c3be8797387d11a3da0edf8b610},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/26958445. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

The devices calibrated most frequently by the acoustical measurement services at the National Institute of Standards and Technology (NIST) over the 50-year period from 1963 to 2012 were one-inch condenser microphones of three specific standard types: LS1Pn, LS1Po, and WS1P. Due to its long history of providing calibrations of such microphones to customers, NIST is in a unique position to analyze data concerning the long-term stability of these devices. This long history has enabled NIST to acquire and aggregate a substantial amount of repeat calibration data for a large number of microphones that belong to various other standards and calibration laboratories. In addition to determining microphone sensitivities at the time of calibration, it is important to have confidence that the microphones do not typically undergo significant drift as compared to the calibration uncertainty during the periods between calibrations. For each of the three microphone types, an average drift rate and approximate 95 % confidence interval were computed by two different statistical methods, and the results from the two methods were found to differ insignificantly in each case. These results apply to typical microphones of these types that are used in a suitable environment and handled with care. The average drift rate for Type LS1Pn microphones was -0.004 dB/year to 0.003 dB/year. The average drift rate for Type LS1Po microphones was -0.016 dB/year to 0.008 dB/year. The average drift rate for Type WS1P microphones was -0.004 dB/year to 0.018 dB/year. For each of these microphone types, the average drift rate is not significantly different from zero. This result is consistent with the performance expected of condenser microphones designed for use as transfer standards. In addition, the values that bound the confidence intervals are well within the limits specified for long-term stability in international standards. Even though these results show very good long-term stability historically for these microphone types, it is expected that periodic calibrations will always be done to track the calibration history of individual microphones and check for anomalies indicative of shifts in sensitivity.

Tema degli esploratori: statistica e scienza dei dati (ricerca Europe PMC «Bayesian statistics», solo open access).
Pubblicazione: Journal of research of the National Institute of Standards and Technology, 2015 (prima pubblicazione 2015-08-17).
Identificativi: PMID 26958445 · PMCID PMC4730669 · DOI 10.6028/jres.120.012.
Parole chiave degli autori: calibration; Condenser Microphone; Long-term Stability; Laboratory Standard Microphone; Microphone Calibration; Working Standard Microphone.
Pagina: https://europepmc.org/article/MED/26958445 — https://doi.org/10.6028/jres.120.012. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0.
Letto il 2026-10-09T22:24:51.645Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.