Capillary drops, capillary pooled, and venous blood samples for determining hemoglobin concentration using HemoCue: A measurement system analysis — Europe PMC PMID:39453907
- Source
- Europe PMC (en)
- Original
- https://europepmc.org/article/MED/39453907
- DOI
- 10.1371/journal.pone.0312233
- License (as archived)
- CC0-1.0
- Archived
- 2026-10-06T15:13:19.073Z
- SHA-256 of the archived text
b8efabb1207f406c6707781607e97edf7fe095d254526213404782e627b28555- Stable URL of this page
https://metraipolis.com/documento/dc_940175eedd78
How to cite
Capillary drops, capillary pooled, and venous blood samples for determining hemoglobin concentration using HemoCue: A measurement system analysis — Europe PMC PMID:39453907 — Europe PMC (en), https://europepmc.org/article/MED/39453907 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_940175eedd78 (Metraipolis archive dc_940175eedd78, sha256 b8efabb1207f406c6707781607e97edf7fe095d254526213404782e627b28555)
BibTeX
@misc{dc940175eedd78,
title = {Capillary drops, capillary pooled, and venous blood samples for determining hemoglobin concentration using HemoCue: A measurement system analysis — Europe PMC PMID:39453907},
howpublished = {Europe PMC (en)},
url = {https://europepmc.org/article/MED/39453907},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_940175eedd78; sha256 b8efabb1207f406c6707781607e97edf7fe095d254526213404782e627b28555},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/39453907. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
Several external and internal factors can affect the performance and variability of Hemoglobin concentration [Hb] measurements using HemoCue, and documentation on the contribution of different sources of [Hb] variation is limited. We used an experimental repeated measurements design with nine randomly selected participants, three HemoCue devices, and three trained field workers. HemoCue measurements for all samples were repeated three times. The [Hb] measurement system considers four sources of error: 1) HemoCue devices, 2) field workers, 3) between individuals, and 4) within individuals. A concordance analysis was used to assess accuracy and precision, and a linear mixed model was used to estimate mean differences (bias) from blood specimens, anticoagulants, and to estimate the contribution of the 4 sources of error to [Hb] measurements. Positive mean [Hb] differences were found: 1.34 g/dL for capillary drops, 0.81 g/dL for pooled capillary blood samples, 0.756 g/dL for venous blood stored with EDTA, and 0.911 g/dL for venous blood stored with heparin. The mean [Hb] difference for venous blood with EDTA was used as a correction factor for all results measured using a HemoCue. After adjustment, capillary drops showed a mean difference of 0.585 g/dL, and pooled capillary samples were not significantly different. The individual variability was 95.8% of total variance, HemoCue devices contributed 2.1% of measurement error, field staff contributed 0.4%, and the residual was 1.7%. The HemoCue [Hb] measurement system is reliable in controlled environments, with a small measurement error of 4.2%. Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera). Pubblicazione: PloS one, 2024 (prima pubblicazione 2024-10-25). Identificativi: PMID 39453907 · PMCID PMC11508077 · DOI 10.1371/journal.pone.0312233. Descrittori MeSH principali: Capillaries; Hemoglobins. Pagina: https://europepmc.org/article/MED/39453907 — https://doi.org/10.1371/journal.pone.0312233. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0. Letto il 2026-10-06T15:11:36.293Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.