Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors — arXiv:2611.03359

Source
arXiv (en)
Original
https://arxiv.org/abs/2611.03359
License (as archived)
CC0-1.0
Archived
2026-11-03T11:56:46.113Z
SHA-256 of the archived text
7347bcbd12604085a73987f6e6ea525af9dc94f38751eef3fba2b7f573b4eb19
Stable URL of this page
https://metraipolis.com/documento/dc_8e7a8d714a10

How to cite

Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors — arXiv:2611.03359 — arXiv (en), https://arxiv.org/abs/2611.03359 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_8e7a8d714a10 (Metraipolis archive dc_8e7a8d714a10, sha256 7347bcbd12604085a73987f6e6ea525af9dc94f38751eef3fba2b7f573b4eb19)
BibTeX
@misc{dc8e7a8d714a10,
  title = {Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors — arXiv:2611.03359},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2611.03359},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_8e7a8d714a10; sha256 7347bcbd12604085a73987f6e6ea525af9dc94f38751eef3fba2b7f573b4eb19},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2611.03359. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Accurate spatial-resolution characterization of flat-panel detectors at megavoltage energies is difficult due to long-tailed cross-talk. Conventional measurements at high energy using line-pair gauges or knife-edge phantoms require expensive, carefully fabricated objects and precise alignment, and despite these, scatter from the phantom, detector, and imaging room can further confound the measurement. We present a method for estimating the detector point spread function (PSF) from the mean noise power spectrum (NPS) calculated from repeated radiographs in air. The approach is mathematically equivalent to an autocovariance-based method but is more practical for long-tailed PSFs. We validate the method using simulated air data and Monte Carlo N-particle (MCNP6) calculations of energy deposition in a flat-panel detector. We show that deconvolving the experimental PSF increases the accuracy of transmission radiographs of step wedges. The method requires no phantom or alignments and is straightforward to reproduce.

Tema degli esploratori: statistica e scienza dei dati (elenco arXiv eess.IV, nuovi lavori del giorno).
Categorie arXiv: Image and Video Processing (eess.IV).
Pagina del lavoro: https://arxiv.org/abs/2611.03359 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina.
Letto il 2026-11-03T11:16:07.695Z dalla pagina https://arxiv.org/list/eess.IV/new (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).