Cross-Device Transfer of a Hospital-Trained Fetal Head Circumference Model to a Low-Cost Handheld Ultrasound Probe — arXiv:2610.40710
- Source
- arXiv (en)
- Original
- https://arxiv.org/abs/2610.40710
- License (as archived)
- CC0-1.0
- Archived
- 2026-11-02T08:14:13.590Z
- SHA-256 of the archived text
f2155948d772eae399ad1522bc808790a0ab2ac704672c6f67feccb33095f577- Stable URL of this page
https://metraipolis.com/documento/dc_74ae4607ff35
How to cite
Cross-Device Transfer of a Hospital-Trained Fetal Head Circumference Model to a Low-Cost Handheld Ultrasound Probe — arXiv:2610.40710 — arXiv (en), https://arxiv.org/abs/2610.40710 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_74ae4607ff35 (Metraipolis archive dc_74ae4607ff35, sha256 f2155948d772eae399ad1522bc808790a0ab2ac704672c6f67feccb33095f577)
BibTeX
@misc{dc74ae4607ff35,
title = {Cross-Device Transfer of a Hospital-Trained Fetal Head Circumference Model to a Low-Cost Handheld Ultrasound Probe — arXiv:2610.40710},
howpublished = {arXiv (en)},
url = {https://arxiv.org/abs/2610.40710},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_74ae4607ff35; sha256 f2155948d772eae399ad1522bc808790a0ab2ac704672c6f67feccb33095f577},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2610.40710. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
Automated fetal head circumference (HC) measurement reaches near-expert accuracy on the HC18 benchmark, but published models are trained and evaluated on hospital-grade equipment. We quantify what happens when such a model is carried to a low-cost handheld probe. A lightweight segmentation network (CSM, 0.13M parameters) was trained on HC18 across five independent runs and applied without modification to images from a $1,350 handheld probe (SonoQue C3), acquired at the same visit as the reference hospital scan in 18 patients. In-domain accuracy was 2.34 +/- 0.43 mm; on handheld images the same model gave 16.12 mm against the hospital record, a 6.9-fold degradation. Input preprocessing, a choice rarely reported, accounted for about a third of that gap: direct resizing gave 49.45 mm against 16.12 mm for an aspect-preserving pipeline. A leave-one-out data-efficiency curve showed that fine-tuning on two images significantly worsened accuracy, and that from four images upward error fell by 0.75 mm per doubling of the local dataset (p = 0.0045), although no training-set size up to 27 images significantly improved on the unmodified model. Much of the residual gap belongs to the reference rather than the model: scored against a manual tracing of the same image the model was already within 10.48 mm, and its disagreement with sonographer calipers on the same frame was a near-constant offset, negative in 10 of 10 cases. A single calibration constant estimated leave-one-out reduced that disagreement from 8.12 mm to 3.49 mm, reaching the intraobserver reproducibility reported for experienced sonographers, but did not improve agreement with the hospital record. Cross-device evaluations should therefore report the input preprocessing used and characterise the precision of their reference standard. Tema degli esploratori: statistica e scienza dei dati (elenco arXiv eess.IV, nuovi lavori del giorno). Categorie arXiv: Image and Video Processing (eess.IV). Pagina del lavoro: https://arxiv.org/abs/2610.40710 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina. Letto il 2026-11-02T07:53:31.861Z dalla pagina https://arxiv.org/list/eess.IV/new (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).