Metraipolis › archive › dc_749c11282164

Efficient physiological control of an integrated system architecture for continuous-flow ventricular assist devices: in-silico study — arXiv:2609.24966

Source
arXiv (en)
Original
https://arxiv.org/abs/2609.24966
License (as archived)
CC0-1.0
Archived
2026-10-10T00:57:39.265Z
SHA-256 of the archived text
c98e598d8d75c50fe997b90a239395fc21df6d497282c5737c51582588937cf3
Stable URL of this page
https://metraipolis.com/documento/dc_749c11282164

How to cite

Efficient physiological control of an integrated system architecture for continuous-flow ventricular assist devices: in-silico study — arXiv:2609.24966 — arXiv (en), https://arxiv.org/abs/2609.24966 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_749c11282164 (Metraipolis archive dc_749c11282164, sha256 c98e598d8d75c50fe997b90a239395fc21df6d497282c5737c51582588937cf3)
BibTeX
@misc{dc749c11282164,
  title = {Efficient physiological control of an integrated system architecture for continuous-flow ventricular assist devices: in-silico study — arXiv:2609.24966},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2609.24966},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_749c11282164; sha256 c98e598d8d75c50fe997b90a239395fc21df6d497282c5737c51582588937cf3},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2609.24966. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

This study presents the development and in silico evaluation of an Integrated System Architecture (ISA) for the physiological control of continuous-flow ventricular assist devices (VADs). The system employs an automatic controller based on pressure measurements at the VAD inflow cannula to estimate heart rate and ventricular filling pressure, enabling dynamic speed adjustments that optimize VAD-patient interaction. The evaluation encompassed 27 simulation scenarios assessing four aspects: operating speed regulation, responsiveness to variable demand, adverse event mitigation, and physiological impact. The controller continuously adapted to preload, afterload, and heart rate, accommodating flow demands with speed adjustments ranging from -10% (-600 rpm) to +43% (+2,600 rpm) relative to the baseline (6,000 rpm). Furthermore, dynamic regulation eliminated ventricular suction and backflow events, which were previously observed in 12 of the 27 uncontrolled scenarios. Hemodynamic and metabolic outcomes demonstrated improved left ventricular ejection fraction (+6.23% to +62.31%), reduced ejection work (-1,658.64 to -206.37 mmHg*mL), decreased pressure-volume loop area (-9,485.17 to -1,056.05 mmHg*mL), and reduced myocardial oxygen consumption (-20.5 to -1.14 mL/min). Concurrently, total oxygen delivery increased (+56.9 to +626.23 mL/min) along with cardiac power (+100 to +3,070 mW). These findings demonstrate that the ISA successfully maintains physiological regulation and mitigates adverse events, establishing a solid foundation for future in vitro validation.

Tema degli esploratori: calcolo e infrastrutture (retro-catalogo arXiv: annunci del 2026-09-22 in eess.SY).
Categorie arXiv: Systems and Control (eess.SY) ; Signal Processing (eess.SP).