Metraipolis › archive › dc_6d480681d5de
Characterization of 3-Dimensional Printing and Casting Materials for use in Computed Tomography and X-ray Imaging Phantoms — Europe PMC PMID:39411105
- Source
- Europe PMC (en)
- Original
- https://europepmc.org/article/MED/39411105
- DOI
- 10.6028/jres.125.029
- License (as archived)
- CC0-1.0
- Archived
- 2026-11-07T12:13:14.069Z
- SHA-256 of the archived text
89abd5e7e1d9024e04403356efac6e1cbd0d0bc77d93d877193a216c96af46df- Stable URL of this page
https://metraipolis.com/documento/dc_6d480681d5de
How to cite
Characterization of 3-Dimensional Printing and Casting Materials for use in Computed Tomography and X-ray Imaging Phantoms — Europe PMC PMID:39411105 — Europe PMC (en), https://europepmc.org/article/MED/39411105 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_6d480681d5de (Metraipolis archive dc_6d480681d5de, sha256 89abd5e7e1d9024e04403356efac6e1cbd0d0bc77d93d877193a216c96af46df)
BibTeX
@misc{dc6d480681d5de,
title = {Characterization of 3-Dimensional Printing and Casting Materials for use in Computed Tomography and X-ray Imaging Phantoms — Europe PMC PMID:39411105},
howpublished = {Europe PMC (en)},
url = {https://europepmc.org/article/MED/39411105},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_6d480681d5de; sha256 89abd5e7e1d9024e04403356efac6e1cbd0d0bc77d93d877193a216c96af46df},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/39411105. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
Imaging phantoms are used to calibrate and validate the performance of medical computed tomography (CT) systems. Many new materials developed for three-dimensional (3D) printing processes may be useful in the direct printing or casting of biomimetic and geometrically accurate CT and X-ray phantoms. The X-ray linear attenuation coefficients of polymer samples were measured to discover materials for use as tissue mimics in phantoms. This study included a cohort of polymer compounds that were tested in cured form. The cohort consisted of 101 standardized polymer samples fabricated from: two-part silicones and polyurethanes used in commercial casting processes; one-part optically cured polyurethanes used in 3D printing; and fused deposition thermoplastics used in 3D printing. The testing was performed with a commercial micro-CT imaging system from 40 kVp to 140 kVp. The X-ray linear coefficients of the samples and human tissues were plotted with error bars to allow the reader to identify suitable mimics. The X-ray linear attenuation coefficients of the tested material samples spanned a wide range of values, with a small number of them overlapping established human tissue mimic values. Twenty 3D printer materials and one castable polyurethane tracked nylon and polymethyl methacrylate (PMMA) as established X-ray mimics for fat. Five 3D printer materials tracked water as an established X-ray mimic for muscle. Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera). Pubblicazione: Journal of research of the National Institute of Standards and Technology, 2020 (prima pubblicazione 2020-09-15). Identificativi: PMID 39411105 · PMCID PMC11441636 · DOI 10.6028/jres.125.029. Parole chiave degli autori: X-ray; Medical imaging; CT; Polymer; computed tomography; Phantom; 3D Printing. Pagina: https://europepmc.org/article/MED/39411105 — https://doi.org/10.6028/jres.125.029. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0. Letto il 2026-11-07T12:08:07.365Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.