Metraipolis › archive › dc_640998b5477a

Wavelength Calibration Accuracy Across the STIS CCD: Pipeline Update and User Guidance — arXiv:2608.10199

Source
arXiv (en)
Original
https://arxiv.org/abs/2608.10199
License (as archived)
CC0-1.0
Archived
2026-10-19T06:41:38.731Z
SHA-256 of the archived text
472431706a94a61241248263b95cf0e29321c9b30c0f574dccd3f69c6b2c1f8a
Stable URL of this page
https://metraipolis.com/documento/dc_640998b5477a

How to cite

Wavelength Calibration Accuracy Across the STIS CCD: Pipeline Update and User Guidance — arXiv:2608.10199 — arXiv (en), https://arxiv.org/abs/2608.10199 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_640998b5477a (Metraipolis archive dc_640998b5477a, sha256 472431706a94a61241248263b95cf0e29321c9b30c0f574dccd3f69c6b2c1f8a)
BibTeX
@misc{dc640998b5477a,
  title = {Wavelength Calibration Accuracy Across the STIS CCD: Pipeline Update and User Guidance — arXiv:2608.10199},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2608.10199},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_640998b5477a; sha256 472431706a94a61241248263b95cf0e29321c9b30c0f574dccd3f69c6b2c1f8a},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2608.10199. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

We investigate the wavelength calibration accuracy of STIS CCD spectra as a function of detector position, focusing on the E1/E2 pseudo-apertures. Comparison between measured lamp line centroids and laboratory wavelengths shows that, while the standard calstis solution is stable at the detector center, significant offsets are present toward the CCD edges and increase with time. We used a test version of calstis in which the wavelength calibration step employs a row-selected cross-correlation, leading to improved wavelength calibration accuracy for edge extractions. Motivated by these results, we implemented an update to calstis4 that applies a row-selected wavecal procedure to observations taken with the E1/E2 pseudo-apertures, while preserving the standard procedure for nominal extractions. The updated pipeline was then used to reprocess the entire STIS archive in MAST. Validation tests confirm improved agreement between nominal and E1/E2 spectra without significant side effects for more than 97% of the datasets in the MAST archive. The main exceptions are the majority of G230MB and a few G230LB and G430M datasets, where the lower signal-to-noise at the CCD edge makes the lamp lines harder to detect and cosmic-ray residuals can dominate the cross-correlation, leading to incorrect wavelength shifts. A Jupyter Notebook is also provided to enable similar corrections in case the cross-correlation fails, for extractions at positions other than E1/E2, or for multiple extractions along the slit. Finally, we discuss the potential impact of the wavelength calibration accuracy on science results.

Tema degli esploratori: fisica e astronomia (retro-catalogo arXiv: annunci del 2026-08-12 in astro-ph.IM).
Categorie arXiv: Instrumentation and Methods for Astrophysics (astro-ph.IM).