Metraipolis › archive › dc_5e234fe90c78

Comparative analysis of three studies measuring fluorescence from engineered bacterial genetic constructs — Europe PMC PMID:34097703

Source
Europe PMC (en)
Original
https://europepmc.org/article/MED/34097703
DOI
10.1371/journal.pone.0252263
License (as archived)
CC0-1.0
Archived
2026-11-04T03:33:29.531Z
SHA-256 of the archived text
1d04fbecf2153b72dd6adcbb7a789e66116d481538313989d69e585e8ab06124
Stable URL of this page
https://metraipolis.com/documento/dc_5e234fe90c78

How to cite

Comparative analysis of three studies measuring fluorescence from engineered bacterial genetic constructs — Europe PMC PMID:34097703 — Europe PMC (en), https://europepmc.org/article/MED/34097703 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_5e234fe90c78 (Metraipolis archive dc_5e234fe90c78, sha256 1d04fbecf2153b72dd6adcbb7a789e66116d481538313989d69e585e8ab06124)
BibTeX
@misc{dc5e234fe90c78,
  title = {Comparative analysis of three studies measuring fluorescence from engineered bacterial genetic constructs — Europe PMC PMID:34097703},
  howpublished = {Europe PMC (en)},
  url = {https://europepmc.org/article/MED/34097703},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_5e234fe90c78; sha256 1d04fbecf2153b72dd6adcbb7a789e66116d481538313989d69e585e8ab06124},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/34097703. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Reproducibility is a key challenge of synthetic biology, but the foundation of reproducibility is only as solid as the reference materials it is built upon. Here we focus on the reproducibility of fluorescence measurements from bacteria transformed with engineered genetic constructs. This comparative analysis comprises three large interlaboratory studies using flow cytometry and plate readers, identical genetic constructs, and compatible unit calibration protocols. Across all three studies, we find similarly high precision in the calibrants used for plate readers. We also find that fluorescence measurements agree closely across the flow cytometry results and two years of plate reader results, with an average standard deviation of 1.52-fold, while the third year of plate reader results are consistently shifted by more than an order of magnitude, with an average shift of 28.9-fold. Analyzing possible sources of error indicates this shift is due to incorrect preparation of the fluorescein calibrant. These findings suggest that measuring fluorescence from engineered constructs is highly reproducible, but also that there is a critical need for access to quality controlled fluorescent calibrants for plate readers.

Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera).
Pubblicazione: PloS one, 2021 (prima pubblicazione 2021-06-07).
Identificativi: PMID 34097703 · PMCID PMC8183995 · DOI 10.1371/journal.pone.0252263.
Pagina: https://europepmc.org/article/MED/34097703 — https://doi.org/10.1371/journal.pone.0252263. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0.
Letto il 2026-11-04T03:26:48.073Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.