Metraipolis › archive › dc_59f09770ef38

RoSE: A Robotic Soft Esophagus for Endoprosthetic Stent Testing — arXiv:2608.09891

Source
arXiv (en)
Original
https://arxiv.org/abs/2608.09891
License (as archived)
CC0-1.0
Archived
2026-10-19T09:12:58.171Z
SHA-256 of the archived text
11e7b9156a9d45b458ce587d2b6ce5d98e487f62fe0e903d85567bce16c11df5
Stable URL of this page
https://metraipolis.com/documento/dc_59f09770ef38

How to cite

RoSE: A Robotic Soft Esophagus for Endoprosthetic Stent Testing — arXiv:2608.09891 — arXiv (en), https://arxiv.org/abs/2608.09891 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_59f09770ef38 (Metraipolis archive dc_59f09770ef38, sha256 11e7b9156a9d45b458ce587d2b6ce5d98e487f62fe0e903d85567bce16c11df5)
BibTeX
@misc{dc59f09770ef38,
  title = {RoSE: A Robotic Soft Esophagus for Endoprosthetic Stent Testing — arXiv:2608.09891},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2608.09891},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_59f09770ef38; sha256 11e7b9156a9d45b458ce587d2b6ce5d98e487f62fe0e903d85567bce16c11df5},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2608.09891. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Soft robotic systems are well suited for developing devices for biomedical applications. A bio-mimicking robotic soft esophagus (RoSE) is developed as an in vitro testing device of endoprosthetic stents for dysphagia management. Endoprosthetic stent placement is an immediate and cost-effective therapy for dysphagia caused by malignant esophageal strictures from esophageal cancer. However, later stage complications, like stent migration, could weaken swallow efficacy in the esophagus. The stent radial force (RF) on the esophageal wall is pivotal in avoiding stent migration. Due to limited randomized controlled trials in patients, stent design and stenting guidelines remain incomplete. To address this knowledge deficit, we investigate RoSE by implanting two stents (A and B) of different radial stiffness characteristics, to measure stent RF and its effect on migration. Endoscopic manometry under peristalsis is also performed to study the impact of stenting and stent dysfunction on intra-bolus pressure signatures (IBPSs) and swallowing efficacy. Each implanted stent undergoes experiments with varied peristalsis velocity, wavelength, and bolus concentrations. The results show that stiffer stent B has a higher RF, whereas stent A maintains a lower RF profile due to lesser stiffness. High RF is necessary to minimize migration under prolonged peristaltic contractions in RoSE. For manometry, stent A slightly increases IBPS, but stiffer stent B significantly decreases IBPS, especially for higher-concentration boluses. If a stiffer stent buckles, it can reduce swallow efficacy and cause recurrent dysphagia. RoSE is therefore an innovative soft robotic platform for testing endoprosthetic stents and addressing clinical challenges in stent evaluation.

Tema degli esploratori: statistica e scienza dei dati (retro-catalogo arXiv: annunci del 2026-08-11 in cs.RO).
Categorie arXiv: Robotics (cs.RO).
Pagina del lavoro: https://arxiv.org/abs/2608.09891 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina.
Letto il 2026-10-19T09:11:58.807Z dalla pagina https://arxiv.org/catchup/cs.RO/2026-08-11?abs=True (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).