Metraipolis › archive › dc_54c876e5915e

AI-based single-shot structured-light depth reconstruction for real-time laparoscopic surgical guidance — arXiv:2608.05109

Source
arXiv (en)
Original
https://arxiv.org/abs/2608.05109
License (as archived)
CC0-1.0
Archived
2026-10-31T11:50:08.242Z
SHA-256 of the archived text
5972333559be6bdaaa7c83552d698e7dc70872196b98fb11afa93ab9bd7ab20c
Stable URL of this page
https://metraipolis.com/documento/dc_54c876e5915e

How to cite

AI-based single-shot structured-light depth reconstruction for real-time laparoscopic surgical guidance — arXiv:2608.05109 — arXiv (en), https://arxiv.org/abs/2608.05109 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_54c876e5915e (Metraipolis archive dc_54c876e5915e, sha256 5972333559be6bdaaa7c83552d698e7dc70872196b98fb11afa93ab9bd7ab20c)
BibTeX
@misc{dc54c876e5915e,
  title = {AI-based single-shot structured-light depth reconstruction for real-time laparoscopic surgical guidance — arXiv:2608.05109},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2608.05109},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_54c876e5915e; sha256 5972333559be6bdaaa7c83552d698e7dc70872196b98fb11afa93ab9bd7ab20c},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2608.05109. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Significance. Accurate intraoperative depth perception is important for autonomous and semi-autonomous robotic laparoscopic surgery. Conventional fringe projection profilometry can achieve millimeter-scale accuracy but often requires multi-shot acquisition, digital-micromirror-device projection, and projector-camera synchronization, complicating integration into compact laparoscopic systems. Aim. To develop a synchronization-free, single-shot depth-sensing platform using a passive LED-illuminated binary mask and a VQ-VAE prior with a custom U-Net depth head. Approach. A compact projection module was coupled to one channel of a dual-channel laparoscope, while the second channel imaged the fringe-illuminated target. A Zivid 3D camera acquired reference depth for 722 paired phantom images. Zivid depth maps were reprojected into the SSLE image frame for supervised training and evaluation. The VQ-VAE encoded each input into a discrete latent representation, and a latent-space U-Net predicted depth without a separate mask-prediction branch. Results. Using a fixed train/validation/test split, the proposed model achieved an MAE of 3.70 mm, AbsRel of 0.0326, delta=1.1 accuracy of 0.962, and delta=1.1^2 accuracy of 0.970. It achieved lower MAE than the dual U-Net MaskNet + DepthNet baseline and outperformed off-the-shelf monocular depth models in MAE, AbsRel, and threshold accuracy. The pipeline operated at 26.0 Hz over 301 consecutive frames on an NVIDIA A100 GPU. Conclusions. The LED-illuminated binary-pattern platform with latent-space depth reconstruction enables synchronization-free, video-rate endoscopic depth estimation. Results demonstrate Zivid-referenced phantom reconstruction without an explicit segmentation stage, while emphasizing the importance of dataset size and SSLE-Zivid calibration accuracy.

Tema degli esploratori: statistica e scienza dei dati (retro-catalogo arXiv: annunci del 2026-08-06 in cs.RO).
Categorie arXiv: Image and Video Processing (eess.IV) ; Robotics (cs.RO); Optics (physics.optics).
Pagina del lavoro: https://arxiv.org/abs/2608.05109 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina.
Letto il 2026-10-31T11:39:29.953Z dalla pagina https://arxiv.org/catchup/cs.RO/2026-08-06?abs=True (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).