Calorimetry in Computed Tomography Beams — Europe PMC PMID:38469437
- Source
- Europe PMC (en)
- Original
- https://europepmc.org/article/MED/38469437
- DOI
- 10.6028/jres.126.054
- License (as archived)
- CC0-1.0
- Archived
- 2026-11-06T12:33:33.196Z
- SHA-256 of the archived text
ce088b74a9a022a72f77304c16acd377066825f30c432ef02bbed167ebf88830- Stable URL of this page
https://metraipolis.com/documento/dc_45be04a61472
How to cite
Calorimetry in Computed Tomography Beams — Europe PMC PMID:38469437 — Europe PMC (en), https://europepmc.org/article/MED/38469437 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_45be04a61472 (Metraipolis archive dc_45be04a61472, sha256 ce088b74a9a022a72f77304c16acd377066825f30c432ef02bbed167ebf88830)
BibTeX
@misc{dc45be04a61472,
title = {Calorimetry in Computed Tomography Beams — Europe PMC PMID:38469437},
howpublished = {Europe PMC (en)},
url = {https://europepmc.org/article/MED/38469437},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_45be04a61472; sha256 ce088b74a9a022a72f77304c16acd377066825f30c432ef02bbed167ebf88830},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/38469437. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
A portable calorimeter for direct realization of absorbed dose in medical computed tomography (CT) procedures was constructed and tested in a positron emission tomography (PET) CT scanner. The calorimeter consists of two small thermistors embedded in a polystyrene (PS) cylindrical "core" (1.5 cm diameter) that can be inserted into a cylindrical high-density polyethylene (HDPE) phantom (30 cm diameter). The cylindrical design of core and phantom allows coaxial alignment of the system with the scanner rotation axis, which is necessary to minimize variations in dose that would otherwise occur as the X-ray source is rotated during scanning operations. The core can be replaced by a cylindrical ionization chamber for comparing dose measurement results. Measurements using the core and a calibrated thimble ionization chamber were carried out in a beam of 6 MV X-rays from a clinical accelerator and in 120 kV X-rays from a CT scanner. Doses obtained from the calorimeter and chamber in the 6 MV beam exhibited good agreement over a range of dose rates from 0.8 Gy/min to 4 Gy/min, with negligible excess heat. For the CT beam, as anticipated for these X-ray energies, the calorimeter response was complicated by excess heat from device components. Analyses done in the frequency domain and time domain indicated that excess heat increased calorimetric temperature rise by a factor of about 15. The calorimeter's response was dominated by dose to the thermistor, which contains high-atomic-number elements. Therefore, for future construction of calorimeters for CT beams, lower-atomic-number temperature sensors will be needed. These results serve as a guide for future alternative design of calorimeters toward a calorimetry absorbed dose standard for diagnostic CT. Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera). Pubblicazione: Journal of research of the National Institute of Standards and Technology, 2021 (prima pubblicazione 2021-01-01). Identificativi: PMID 38469437 · PMCID PMC10046824 · DOI 10.6028/jres.126.054. Parole chiave degli autori: Absorbed dose; Polystyrene; Calorimeter; Thermistor; Ct Dose; Medium Energy X-rays. Pagina: https://europepmc.org/article/MED/38469437 — https://doi.org/10.6028/jres.126.054. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0. Letto il 2026-11-06T12:30:35.636Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.