Accelerating whole-heart 3D T2 mapping: Impact of undersampling strategies and reconstruction techniques — Europe PMC PMID:34506496
- Source
- Europe PMC (en)
- Original
- https://europepmc.org/article/MED/34506496
- DOI
- 10.1371/journal.pone.0252777
- License (as archived)
- CC0-1.0
- Archived
- 2026-11-02T22:26:53.997Z
- SHA-256 of the archived text
0c1ce09c1ce5d7eda209052c3a4d99eab1a88ce35705eb8d7d9b9877ab388422- Stable URL of this page
https://metraipolis.com/documento/dc_3df89dac1a60
How to cite
Accelerating whole-heart 3D T2 mapping: Impact of undersampling strategies and reconstruction techniques — Europe PMC PMID:34506496 — Europe PMC (en), https://europepmc.org/article/MED/34506496 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_3df89dac1a60 (Metraipolis archive dc_3df89dac1a60, sha256 0c1ce09c1ce5d7eda209052c3a4d99eab1a88ce35705eb8d7d9b9877ab388422)
BibTeX
@misc{dc3df89dac1a60,
title = {Accelerating whole-heart 3D T2 mapping: Impact of undersampling strategies and reconstruction techniques — Europe PMC PMID:34506496},
howpublished = {Europe PMC (en)},
url = {https://europepmc.org/article/MED/34506496},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_3df89dac1a60; sha256 0c1ce09c1ce5d7eda209052c3a4d99eab1a88ce35705eb8d7d9b9877ab388422},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/34506496. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
Purpose: We aim to determine an advantageous approach for the acceleration of high spatial resolution 3D cardiac T2 relaxometry data by comparing the performance of different undersampling patterns and reconstruction methods over a range of acceleration rates. Methods: Multi-volume 3D high-resolution cardiac images were acquired fully and undersampled retrospectively using 1) optimal CAIPIRINHA and 2) a variable density random (VDR) sampling. Data were reconstructed using 1) multi-volume sensitivity encoding (SENSE), 2) joint-sparsity SENSE and 3) model-based SENSE. Four metrics were calculated on 3 naïve swine and 8 normal human subjects over a whole left-ventricular region of interest: root-mean-square error (RMSE) of image signal intensity, RMSE of T2, the bias of mean T2, and standard deviation (SD) of T2. Fully sampled data and volume-by-volume SENSE with standard equally spaced undersampling were used as references. The Jaccard index calculated from one swine with acute myocardial infarction (MI) was used to demonstrate preservation of segmentation of edematous tissues with elevated T2. Results: In naïve swine and normal human subjects, all methods had similar performance when the net reduction factor (Rnet) 2.5, while VDR sampling with the joint-sparsity SENSE had the lowest bias of mean T2 (0.0-1.1ms) when Rnet>3. The RMSEs of parametric T2 values (9.2%-24.6%) were larger than for image signal intensities (5.2%-18.4%). In the swine with MI, VDR sampling with either joint-sparsity or model-based SENSE showed consistently higher Jaccard index for all Rnet (0.71-0.50) than volume-by-volume SENSE (0.68-0.30). Conclusions: Retrospective exploration of undersampling and reconstruction in 3D whole-heart T2 parametric mapping revealed that maps were more sensitive to undersampling than images, presenting a more stringent limiting factor on Rnet. The combination of VDR sampling patterns with model-based or joint-sparsity SENSE reconstructions were more robust for Rnet>3. Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera). Pubblicazione: PloS one, 2021 (prima pubblicazione 2021-09-10). Identificativi: PMID 34506496 · PMCID PMC8432823 · DOI 10.1371/journal.pone.0252777. Pagina: https://europepmc.org/article/MED/34506496 — https://doi.org/10.1371/journal.pone.0252777. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0. Letto il 2026-11-02T22:18:39.948Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.