Metraipolis › archive › dc_3657a77611bf

Tolerance-Dependent Inspection Disagreement Between a Fixed CMM and a Portable Articulated-Arm CMM — arXiv:2608.21404

Source
arXiv (en)
Original
https://arxiv.org/abs/2608.21404
License (as archived)
CC0-1.0
Archived
2026-10-18T09:01:42.305Z
SHA-256 of the archived text
e211ed302b0cdbde82ccf547d4b4f4bc676be62eb09edd5b5ae4e8bb04841129
Stable URL of this page
https://metraipolis.com/documento/dc_3657a77611bf

How to cite

Tolerance-Dependent Inspection Disagreement Between a Fixed CMM and a Portable Articulated-Arm CMM — arXiv:2608.21404 — arXiv (en), https://arxiv.org/abs/2608.21404 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_3657a77611bf (Metraipolis archive dc_3657a77611bf, sha256 e211ed302b0cdbde82ccf547d4b4f4bc676be62eb09edd5b5ae4e8bb04841129)
BibTeX
@misc{dc3657a77611bf,
  title = {Tolerance-Dependent Inspection Disagreement Between a Fixed CMM and a Portable Articulated-Arm CMM — arXiv:2608.21404},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2608.21404},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_3657a77611bf; sha256 e211ed302b0cdbde82ccf547d4b4f4bc676be62eb09edd5b5ae4e8bb04841129},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2608.21404. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Fixed coordinate measuring machines (CMMs) and portable articulated-arm CMMs are often assigned to the same inspection task, but their nominal accuracy specifications do not show whether a change of instrument will preserve the disposition of a part. The question is not simply how far the two results differ, but whether that difference crosses the tolerance boundary. We examined this issue with recorded measurements of cylindrical, cubic, and spherical features under nominal 20 °C and 30 °C conditions. Repeated records and two roughness profiles without sufficient acquisition information were removed, leaving six dimensional and four form profiles. For each dimensional feature, the distances of the two system means from nominal define the exact tolerance interval in which the systems receive opposite direct labels. The fixed-CMM stream was approximately 11.2 {\mu}m higher than the articulated-arm stream at both conditions. All four form profiles fell on opposite sides of the recorded 10 {\mu}m upper limit. The dimensional disagreement intervals also overlapped strongly; their mean widths were 6.573 {\mu}m at 20 °C and 4.995 {\mu}m at 30 °C. The results clarify why an average difference between instruments is not, by itself, a measure of substitution risk. The proposed tolerance map identifies the feature-tolerance combinations for which instrument choice can change the recorded inspection label and, therefore, where a controlled equivalence study and a task-specific uncertainty budget are needed before substitution.

Tema degli esploratori: statistica e scienza dei dati (retro-catalogo arXiv: annunci del 2026-08-25 in cs.RO).
Categorie arXiv: Robotics (cs.RO) ; Signal Processing (eess.SP).
Pagina del lavoro: https://arxiv.org/abs/2608.21404 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina.
Letto il 2026-10-18T08:57:47.824Z dalla pagina https://arxiv.org/catchup/cs.RO/2026-08-25?abs=True (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).