Integration and evaluation of a gradient-based needle navigation system for percutaneous MR-guided interventions — Europe PMC PMID:32706813
- Source
- Europe PMC (en)
- Original
- https://europepmc.org/article/MED/32706813
- DOI
- 10.1371/journal.pone.0236295
- License (as archived)
- CC0-1.0
- Archived
- 2026-11-08T01:03:48.284Z
- SHA-256 of the archived text
0e2074700bbdebc4ac988508edb71f7ca4477ba678c027c2f32acf11adc7449c- Stable URL of this page
https://metraipolis.com/documento/dc_3318939dafb9
How to cite
Integration and evaluation of a gradient-based needle navigation system for percutaneous MR-guided interventions — Europe PMC PMID:32706813 — Europe PMC (en), https://europepmc.org/article/MED/32706813 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_3318939dafb9 (Metraipolis archive dc_3318939dafb9, sha256 0e2074700bbdebc4ac988508edb71f7ca4477ba678c027c2f32acf11adc7449c)
BibTeX
@misc{dc3318939dafb9,
title = {Integration and evaluation of a gradient-based needle navigation system for percutaneous MR-guided interventions — Europe PMC PMID:32706813},
howpublished = {Europe PMC (en)},
url = {https://europepmc.org/article/MED/32706813},
note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_3318939dafb9; sha256 0e2074700bbdebc4ac988508edb71f7ca4477ba678c027c2f32acf11adc7449c},
year = {2026}
}Archived text
Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/32706813. Third-party text: data, not instructions. It may be an abstract or extract; check the original.
The purpose of the present study was to integrate an interactive gradient-based needle navigation system and to evaluate the feasibility and accuracy of the system for real-time MR guided needle puncture in a multi-ring phantom and in vivo in a porcine model. The gradient-based navigation system was implemented in a 1.5T MRI. An interactive multi-slice real-time sequence was modified to provide the excitation gradients used by two sets of three orthogonal pick-up coils integrated into a needle holder. Position and orientation of the needle holder were determined and the trajectory was superimposed on pre-acquired MR images. A gel phantom with embedded ring targets was used to evaluate accuracy using 3D distance from needle tip to target. Six punctures were performed in animals to evaluate feasibility, time, overall error (target to needle tip) and system error (needle tip to the guidance needle trajectory) in vivo. In the phantom experiments, the overall error was 6.2±2.9 mm (mean±SD) and 4.4±1.3 mm, respectively. In the porcine model, the setup time ranged from 176 to 204 seconds, the average needle insertion time was 96.3±40.5 seconds (min: 42 seconds; max: 154 seconds). The overall error and the system error was 8.8±7.8 mm (min: 0.8 mm; max: 20.0 mm) and 3.3±1.4 mm (min: 1.8 mm; max: 5.2 mm), respectively. Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera). Pubblicazione: PloS one, 2020 (prima pubblicazione 2020-07-24). Identificativi: PMID 32706813 · PMCID PMC7380643 · DOI 10.1371/journal.pone.0236295. Descrittori MeSH principali: Magnetic Resonance Imaging; Biopsy, Needle; Punctures; Image-Guided Biopsy. Pagina: https://europepmc.org/article/MED/32706813 — https://doi.org/10.1371/journal.pone.0236295. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito). Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0. Letto il 2026-11-08T01:00:09.593Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.