Multi-site, multi-platform comparison of MRI T1 measurement using the system phantom — Europe PMC PMID:34191819

Source
Europe PMC (en)
Original
https://europepmc.org/article/MED/34191819
DOI
10.1371/journal.pone.0252966
License (as archived)
CC0-1.0
Archived
2026-11-03T21:35:32.267Z
SHA-256 of the archived text
ae2528a9451561dcf700631fc67307c527435f297584d456d41a6558dfc960d7
Stable URL of this page
https://metraipolis.com/documento/dc_2fd6f67f965e

How to cite

Multi-site, multi-platform comparison of MRI T1 measurement using the system phantom — Europe PMC PMID:34191819 — Europe PMC (en), https://europepmc.org/article/MED/34191819 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_2fd6f67f965e (Metraipolis archive dc_2fd6f67f965e, sha256 ae2528a9451561dcf700631fc67307c527435f297584d456d41a6558dfc960d7)
BibTeX
@misc{dc2fd6f67f965e,
  title = {Multi-site, multi-platform comparison of MRI T1 measurement using the system phantom — Europe PMC PMID:34191819},
  howpublished = {Europe PMC (en)},
  url = {https://europepmc.org/article/MED/34191819},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_2fd6f67f965e; sha256 ae2528a9451561dcf700631fc67307c527435f297584d456d41a6558dfc960d7},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/34191819. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Recent innovations in quantitative magnetic resonance imaging (MRI) measurement methods have led to improvements in accuracy, repeatability, and acquisition speed, and have prompted renewed interest to reevaluate the medical value of quantitative T1. The purpose of this study was to determine the bias and reproducibility of T1 measurements in a variety of MRI systems with an eye toward assessing the feasibility of applying diagnostic threshold T1 measurement across multiple clinical sites. We used the International Society of Magnetic Resonance in Medicine/National Institute of Standards and Technology (ISMRM/NIST) system phantom to assess variations of T1 measurements, using a slow, reference standard inversion recovery sequence and a rapid, commonly-available variable flip angle sequence, across MRI systems at 1.5 tesla (T) (two vendors, with number of MRI systems n = 9) and 3 T (three vendors, n = 18). We compared the T1 measurements from inversion recovery and variable flip angle scans to ISMRM/NIST phantom reference values using Analysis of Variance (ANOVA) to test for statistical differences between T1 measurements grouped according to MRI scanner manufacturers and/or static field strengths. The inversion recovery method had minor over- and under-estimations compared to the NMR-measured T1 values at both 1.5 T and 3 T. Variable flip angle measurements had substantially greater deviations from the NMR-measured T1 values than the inversion recovery measurements. At 3 T, the measured variable flip angle T1 for one vendor is significantly different than the other two vendors for most of the samples throughout the clinically relevant range of T1. There was no consistent pattern of discrepancy between vendors. We suggest establishing rigorous quality control procedures for validating quantitative MRI methods to promote confidence and stability in associated measurement techniques and to enable translation of diagnostic threshold from the research center to the entire clinical community.

Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera).
Pubblicazione: PloS one, 2021 (prima pubblicazione 2021-06-30).
Identificativi: PMID 34191819 · PMCID PMC8244851 · DOI 10.1371/journal.pone.0252966.
Descrittori MeSH principali: Phantoms, Imaging.
Pagina: https://europepmc.org/article/MED/34191819 — https://doi.org/10.1371/journal.pone.0252966. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0.
Letto il 2026-11-03T21:27:34.284Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.