Metraipolis › archive › dc_292154ee6798

Design and Validation of a Lightweight, Low-Profile Powered Knee Prosthesis with Quasi-Direct Drive Actuation — arXiv:2609.02003

Source
arXiv (en)
Original
https://arxiv.org/abs/2609.02003
License (as archived)
CC0-1.0
Archived
2026-10-17T05:51:11.373Z
SHA-256 of the archived text
9cd684891f88890d5c29d18b31aa033ca9490dd2d6bff107fdb1e4f0b0c6136a
Stable URL of this page
https://metraipolis.com/documento/dc_292154ee6798

How to cite

Design and Validation of a Lightweight, Low-Profile Powered Knee Prosthesis with Quasi-Direct Drive Actuation — arXiv:2609.02003 — arXiv (en), https://arxiv.org/abs/2609.02003 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_292154ee6798 (Metraipolis archive dc_292154ee6798, sha256 9cd684891f88890d5c29d18b31aa033ca9490dd2d6bff107fdb1e4f0b0c6136a)
BibTeX
@misc{dc292154ee6798,
  title = {Design and Validation of a Lightweight, Low-Profile Powered Knee Prosthesis with Quasi-Direct Drive Actuation — arXiv:2609.02003},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2609.02003},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_292154ee6798; sha256 9cd684891f88890d5c29d18b31aa033ca9490dd2d6bff107fdb1e4f0b0c6136a},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2609.02003. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Fully-powered knee prostheses, unlike traditional passive knees, can perform controlled positive work, reducing the need for compensatory behaviors by users during energy-intensive activities. While quasi-direct drive (QDD) actuators provide superior torque control, backdrivability, and acoustic noise properties compared to traditional highly-geared actuators, prior QDD prototypes have been too heavy and bulky for commercial translation. In this work, we present the design and validation of a new lightweight (2.6 kg) and low-profile (24.5 cm tip-to-tip build height) QDD knee prosthesis. By optimizing an 18 to 1 two-stage transmission alongside thermal and structural finite-element analyses, we significantly reduce device mass while enabling a peak torque of 145 Nm. Through benchtop tests, we validate the device's high output torque, low backdrive torque (1 Nm), and its precision position and torque control capabilities. We also demonstrate biomimetic kinematics and peak knee extension torques (within one standard deviation of able-bodied references) during both level-ground walking and sit-stand transitions performed by three participants with transfemoral amputation and varying K-levels. By meeting or improving upon the mass, build height, peak torque, and acoustic noise of a leading commercial powered knee, this work establishes the clinical viability of emerging QDD prostheses that promise improved dynamic performance for their users.

Tema degli esploratori: statistica e scienza dei dati (retro-catalogo arXiv: annunci del 2026-09-03 in cs.RO).
Categorie arXiv: Robotics (cs.RO).
Pagina del lavoro: https://arxiv.org/abs/2609.02003 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina.
Letto il 2026-10-17T05:39:02.669Z dalla pagina https://arxiv.org/catchup/cs.RO/2026-09-03?abs=True (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).