Metraipolis › archive › dc_167d367b2c4b

Effect of thermal ballast loading on temperature stability of domestic refrigerators used for vaccine storage — Europe PMC PMID:32639973

Source
Europe PMC (en)
Original
https://europepmc.org/article/MED/32639973
DOI
10.1371/journal.pone.0235777
License (as archived)
CC0-1.0
Archived
2026-11-08T06:11:02.703Z
SHA-256 of the archived text
666effc20d99bb760663b629fe9cebb4d94c2b6a109b30e2e13fcf1a1b6b2ac3
Stable URL of this page
https://metraipolis.com/documento/dc_167d367b2c4b

How to cite

Effect of thermal ballast loading on temperature stability of domestic refrigerators used for vaccine storage — Europe PMC PMID:32639973 — Europe PMC (en), https://europepmc.org/article/MED/32639973 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_167d367b2c4b (Metraipolis archive dc_167d367b2c4b, sha256 666effc20d99bb760663b629fe9cebb4d94c2b6a109b30e2e13fcf1a1b6b2ac3)
BibTeX
@misc{dc167d367b2c4b,
  title = {Effect of thermal ballast loading on temperature stability of domestic refrigerators used for vaccine storage — Europe PMC PMID:32639973},
  howpublished = {Europe PMC (en)},
  url = {https://europepmc.org/article/MED/32639973},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_167d367b2c4b; sha256 666effc20d99bb760663b629fe9cebb4d94c2b6a109b30e2e13fcf1a1b6b2ac3},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/32639973. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Vaccine temperature control failures represent a significant public and private healthcare cost. Vaccines damaged by excessive heat or freezing lose their effectiveness, putting public health at risk. Some vaccine administration programs recommend placing water bottles inside domestic refrigerators used for vaccine storage as a thermal ballast, to mitigate temperature excursion risks. However, the effect of variable thermal ballast loading on refrigerator performance has not been thoroughly quantified or documented, and generalized programmatic recommendations are subject to end-user interpretation. Here we show that a thermal ballast load comprising ten to fifteen percent of the total refrigerator storage volume provides a measurable effect on domestic refrigerator temperature stability during power outage events, maintaining vaccine temperatures between 2 °C and 8 °C for 4 to 6 hours without power. Thermal ballast usage does not reliably reduce the frequency or severity of temperature excursions caused by repeated door opening, accidental "door left open" events, or refrigerator defrost cycle activation. Use of a moderate thermal ballast load is a practical strategy for mitigating temperature excursions risks in areas with frequent or protracted power outages, but the practice has limited benefit in other adverse scenarios. Empowering providers to make informed decisions about the use of thermal ballast materials supports better, safer vaccine management.

Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera).
Pubblicazione: PloS one, 2020 (prima pubblicazione 2020-07-08).
Identificativi: PMID 32639973 · PMCID PMC7343171 · DOI 10.1371/journal.pone.0235777.
Descrittori MeSH principali: Drug Storage.
Pagina: https://europepmc.org/article/MED/32639973 — https://doi.org/10.1371/journal.pone.0235777. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0.
Letto il 2026-11-08T06:07:37.900Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.