Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors — arXiv:2611.03359

Source
arXiv (en)
Original
https://arxiv.org/abs/2611.03359
License (as archived)
CC0-1.0
Archived
2026-11-03T19:10:58.336Z
SHA-256 of the archived text
a147d2ec17eaa5962954d17a70897421e206acc536562c5bce7cc67f1483c530
Stable URL of this page
https://metraipolis.com/documento/dc_166c46ff8883

How to cite

Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors — arXiv:2611.03359 — arXiv (en), https://arxiv.org/abs/2611.03359 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_166c46ff8883 (Metraipolis archive dc_166c46ff8883, sha256 a147d2ec17eaa5962954d17a70897421e206acc536562c5bce7cc67f1483c530)
BibTeX
@misc{dc166c46ff8883,
  title = {Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors — arXiv:2611.03359},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2611.03359},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_166c46ff8883; sha256 a147d2ec17eaa5962954d17a70897421e206acc536562c5bce7cc67f1483c530},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2611.03359. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Accurate spatial-resolution characterization of flat-panel detectors at megavoltage energies is difficult due to long-tailed cross-talk. Conventional measurements at high energy using line-pair gauges or knife-edge phantoms require expensive, carefully fabricated objects and precise alignment, and despite these, scatter from the phantom, detector, and imaging room can further confound the measurement. We present a method for estimating the detector point spread function (PSF) from the mean noise power spectrum (NPS) calculated from repeated radiographs in air. The approach is mathematically equivalent to an autocovariance-based method but is more practical for long-tailed PSFs. We validate the method using simulated air data and Monte Carlo N-particle (MCNP6) calculations of energy deposition in a flat-panel detector. We show that deconvolving the experimental PSF increases the accuracy of transmission radiographs of step wedges. The method requires no phantom or alignments and is straightforward to reproduce.

Tema degli esploratori: statistica e scienza dei dati (elenco arXiv eess.IV, nuovi lavori del giorno).
Categorie arXiv: Image and Video Processing (eess.IV).
Pagina del lavoro: https://arxiv.org/abs/2611.03359 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina.
Letto il 2026-11-03T18:26:22.770Z dalla pagina https://arxiv.org/list/eess.IV/new (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).