Scan-Coil Delay Causes Anisotropic Signal Loss in Fast 4D-STEM — arXiv:2608.13106

Source
arXiv (en)
Original
https://arxiv.org/abs/2608.13106
License (as archived)
CC0-1.0
Archived
2026-10-19T05:11:56.707Z
SHA-256 of the archived text
94d4711e0dfba9b96b41a80eb991c5fbacca6c0195f19b2525d453ea5215a469
Stable URL of this page
https://metraipolis.com/documento/dc_11ada57dd02b

How to cite

Scan-Coil Delay Causes Anisotropic Signal Loss in Fast 4D-STEM — arXiv:2608.13106 — arXiv (en), https://arxiv.org/abs/2608.13106 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_11ada57dd02b (Metraipolis archive dc_11ada57dd02b, sha256 94d4711e0dfba9b96b41a80eb991c5fbacca6c0195f19b2525d453ea5215a469)
BibTeX
@misc{dc11ada57dd02b,
  title = {Scan-Coil Delay Causes Anisotropic Signal Loss in Fast 4D-STEM — arXiv:2608.13106},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2608.13106},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_11ada57dd02b; sha256 94d4711e0dfba9b96b41a80eb991c5fbacca6c0195f19b2525d453ea5215a469},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2608.13106. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Fast pixelated detectors are driving 4D-STEM toward microsecond dwell times, a regime in which the finite response of the scan deflection coils becomes comparable to the dwell time itself. Using direct probe imaging and sub-frame diffraction analysis, we document a significant intra-dwell scan-coil delay that systematically smears the recorded signal anisotropically along the fast scan direction, with a settling timescale of several tens of microseconds. We present a phase-correlation-based sub-frame alignment procedure that measures and corrects this smearing, and we assess its impact on focused and defocused 4D-STEM reconstructions over a range of scan step sizes. The correction restores signal across a broad range of spatial frequencies, with the largest gains at the large step sizes required for low-dose biological imaging. Because it operates on existing data with no modification to the microscope, the method offers a practical route to recovering signal that would otherwise be lost to scan-coil delay.

Tema degli esploratori: scienze della vita e salute (retro-catalogo arXiv: annunci del 2026-08-14 in q-bio.QM).
Categorie arXiv: Instrumentation and Detectors (physics.ins-det) ; Biological Physics (physics.bio-ph); Medical Physics (physics.med-ph); Optics (physics.optics); Quantitative Methods (q-bio.QM).
Pagina del lavoro: https://arxiv.org/abs/2608.13106 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).