Metraipolis › archive › dc_11a3386ba609

The NIST Vacuum Double-Crystal Spectrometer: A Tool for SI-Traceable Measurement of X-Ray Emission Spectra — Europe PMC PMID:38469442

Source
Europe PMC (en)
Original
https://europepmc.org/article/MED/38469442
DOI
10.6028/jres.126.049
License (as archived)
CC0-1.0
Archived
2026-11-06T11:43:20.958Z
SHA-256 of the archived text
c20e3a6ff253c2453768577ab0c3528036b56765d64cd6d0d2078f801542bf42
Stable URL of this page
https://metraipolis.com/documento/dc_11a3386ba609

How to cite

The NIST Vacuum Double-Crystal Spectrometer: A Tool for SI-Traceable Measurement of X-Ray Emission Spectra — Europe PMC PMID:38469442 — Europe PMC (en), https://europepmc.org/article/MED/38469442 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_11a3386ba609 (Metraipolis archive dc_11a3386ba609, sha256 c20e3a6ff253c2453768577ab0c3528036b56765d64cd6d0d2078f801542bf42)
BibTeX
@misc{dc11a3386ba609,
  title = {The NIST Vacuum Double-Crystal Spectrometer: A Tool for SI-Traceable Measurement of X-Ray Emission Spectra — Europe PMC PMID:38469442},
  howpublished = {Europe PMC (en)},
  url = {https://europepmc.org/article/MED/38469442},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_11a3386ba609; sha256 c20e3a6ff253c2453768577ab0c3528036b56765d64cd6d0d2078f801542bf42},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/38469442. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

The NIST Vacuum Double-Crystal Spectrometer (VDCS) has been modernized and is now capable of recording reference-free wavelength-dispersive spectra in the 2 keV to 12 keV x-ray energy range. The VDCS employs crystals in which the lattice spacings are traceable to the definition of the meter through x-ray optical interferometry with a relative uncertainty ﹤10-⁸. VDCS wavelength determination relies upon precision angle difference measurements for which the encoders of the rotation stages have been calibrated using the circle closure method for accurate, absolute angle measurement. The new vacuum-compatible area detector allows quantification of the aberration functions contributing to the observed line shape and in situ alignment of the crystal optics. This latter procedure is augmented with the use of a thin lamella as the first crystal. With these new techniques, x-ray spectra are registered with the VDCS on an absolute energy scale with a relative uncertainty of 10-⁶.

Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera).
Pubblicazione: Journal of research of the National Institute of Standards and Technology, 2021 (prima pubblicazione 2021-01-01).
Identificativi: PMID 38469442 · PMCID PMC10046760 · DOI 10.6028/jres.126.049.
Parole chiave degli autori: X-ray Spectroscopy; Wavelength Standards; Double-crystal Spectrometer; X-ray Transition Energies.
Pagina: https://europepmc.org/article/MED/38469442 — https://doi.org/10.6028/jres.126.049. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0.
Letto il 2026-11-06T11:40:00.691Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.