Closed-loop control of k-space sampling via physiologic feedback for cine MRI — Europe PMC PMID:33373391

Source
Europe PMC (en)
Original
https://europepmc.org/article/MED/33373391
DOI
10.1371/journal.pone.0244286
License (as archived)
CC0-1.0
Archived
2026-11-06T12:37:33.895Z
SHA-256 of the archived text
54ff37617fd28e5cb75b64cdbdbac67d15d29c1c707d2e8a2e9e2ac7563fd816
Stable URL of this page
https://metraipolis.com/documento/dc_0cc563de963b

How to cite

Closed-loop control of k-space sampling via physiologic feedback for cine MRI — Europe PMC PMID:33373391 — Europe PMC (en), https://europepmc.org/article/MED/33373391 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_0cc563de963b (Metraipolis archive dc_0cc563de963b, sha256 54ff37617fd28e5cb75b64cdbdbac67d15d29c1c707d2e8a2e9e2ac7563fd816)
BibTeX
@misc{dc0cc563de963b,
  title = {Closed-loop control of k-space sampling via physiologic feedback for cine MRI — Europe PMC PMID:33373391},
  howpublished = {Europe PMC (en)},
  url = {https://europepmc.org/article/MED/33373391},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_0cc563de963b; sha256 54ff37617fd28e5cb75b64cdbdbac67d15d29c1c707d2e8a2e9e2ac7563fd816},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/33373391. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Background: Segmented cine cardiac MRI combines data from multiple heartbeats to achieve high spatiotemporal resolution cardiac images, yet predefined k-space segmentation trajectories can lead to suboptimal k-space sampling. In this work, we developed and evaluated an autonomous and closed-loop control system for radial k-space sampling (ARKS) to increase sampling uniformity. Methods: The closed-loop system autonomously selects radial k-space sampling trajectory during live segmented cine MRI and attempts to optimize angular sampling uniformity by selecting views in regions of k-space that were not previously well-sampled. Sampling uniformity and the ability to detect cardiac phase in vivo was assessed using ECG data acquired from 10 normal subjects in an MRI scanner. The approach was then implemented with a fast gradient echo sequence on a whole-body clinical MRI scanner and imaging was performed in 4 healthy volunteers. The closed-loop k-space trajectory was compared to random, uniformly distributed and golden angle view trajectories via measurement of k-space uniformity and the point spread function. Lastly, an arrhythmic dataset was used to evaluate a potential application of the approach. Results: The autonomous trajectory increased k-space sampling uniformity by 15±7%, main lobe point spread function (PSF) signal intensity by 6±4%, and reduced ringing relative to golden angle sampling. When implemented, the autonomous pulse sequence prescribed radial view angles faster than the scan TR (0.98 ± 0.01 ms, maximum = 1.38 ms) and increased k-space sampling mean uniformity by 10±11%, decreased uniformity variability by 44±12%, and increased PSF signal ratio by 6±6% relative to golden angle sampling. Conclusion: The closed-loop approach enables near-uniform radial sampling in a segmented acquisition approach which was higher than predetermined golden-angle radial sampling. This can be utilized to increase the sampling or decrease the temporal footprint of an acquisition and the closed-loop framework has the potential to be applied to patients with complex heart rhythms.

Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera).
Pubblicazione: PloS one, 2020 (prima pubblicazione 2020-12-29).
Identificativi: PMID 33373391 · PMCID PMC7771662 · DOI 10.1371/journal.pone.0244286.
Pagina: https://europepmc.org/article/MED/33373391 — https://doi.org/10.1371/journal.pone.0244286. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0.
Letto il 2026-11-06T12:32:34.109Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.