Metraipolis › archive › dc_0b119c990942

Contact-Adaptive Robotic Ultrasound Probe Control for Tissue Exploration and Continuous Task-Relevant Visualization Using Robot-Free Image-Motion Demonstration — arXiv:2609.37456

Source
arXiv (en)
Original
https://arxiv.org/abs/2609.37456
License (as archived)
CC0-1.0
Archived
2026-10-07T19:15:40.203Z
SHA-256 of the archived text
1f015239f68ea8b92248cdb1102bb51957b3eef5689e610c42bb0184f4dc8334
Stable URL of this page
https://metraipolis.com/documento/dc_0b119c990942

How to cite

Contact-Adaptive Robotic Ultrasound Probe Control for Tissue Exploration and Continuous Task-Relevant Visualization Using Robot-Free Image-Motion Demonstration — arXiv:2609.37456 — arXiv (en), https://arxiv.org/abs/2609.37456 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_0b119c990942 (Metraipolis archive dc_0b119c990942, sha256 1f015239f68ea8b92248cdb1102bb51957b3eef5689e610c42bb0184f4dc8334)
BibTeX
@misc{dc0b119c990942,
  title = {Contact-Adaptive Robotic Ultrasound Probe Control for Tissue Exploration and Continuous Task-Relevant Visualization Using Robot-Free Image-Motion Demonstration — arXiv:2609.37456},
  howpublished = {arXiv (en)},
  url = {https://arxiv.org/abs/2609.37456},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_0b119c990942; sha256 1f015239f68ea8b92248cdb1102bb51957b3eef5689e610c42bb0184f4dc8334},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: arXiv (en), https://arxiv.org/abs/2609.37456. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

Robotic ultrasound commonly targets standardized views, predefined scanning protocols, or expert-scan reproduction. We target a different role: while a clinician performs the primary procedure, a robotic assistant maintains a clinician-selected view so that changing tissue remains observable. We present contact-adaptive robotic ultrasound monitoring derived from robot-free demonstrations. Sonologger clamps a six-axis inertial sensor to the clinician-operated probe and records synchronized ultrasound and probe-frame relative rotations without a robot, camera, or tracker. Rather than imitating trajectories, the recordings measure which probe axis and sign the clinician uses to correct an image offset. A measured image latency is removed before still-move-still segmentation yields relative-rotation labels; translation is excluded from control because of inertial drift. These measurements parameterize the Demonstration-Informed Image-Motion Controller (DIMC), an interpretable, non-learned controller combined with beam-axis force regulation and an independent safety supervisor with 5 N forced retreat. On a dynamic bladder phantom, DIMC acquired and held a task-relevant view in 7 of 7 blind, pose-paired trials versus 1 of 7 for a static hold (McNemar p = 0.031), and retained the view during syringe-driven hydro-distension while the supervisor bounded the force excursion. The framework offers an operational representation of ultrasound context for contact-adaptive intraoperative monitoring; clinical validation remains future work.

Tema degli esploratori: statistica e scienza dei dati (retro-catalogo arXiv: annunci del 2026-09-30 in cs.RO).
Categorie arXiv: Robotics (cs.RO).
Pagina del lavoro: https://arxiv.org/abs/2609.37456 — autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza: titolo e abstract sono metadati arXiv distribuiti in CC0 1.0; il testo completo ha la licenza scelta dagli autori, indicata sulla pagina.
Letto il 2026-10-07T18:40:59.512Z dalla pagina https://arxiv.org/catchup/cs.RO/2026-09-30?abs=True (consentita dal robots.txt di arxiv.org, con il suo crawl-delay).