Metraipolis › archive › dc_0889ae59cd8a

X-Ray Computed Tomography Instrument Performance Evaluation, Part III: Sensitivity to Detector Geometry and Rotation Stage Errors at Different Magnifications — Europe PMC PMID:38469439

Source
Europe PMC (en)
Original
https://europepmc.org/article/MED/38469439
DOI
10.6028/jres.126.029
License (as archived)
CC0-1.0
Archived
2026-11-02T15:16:52.868Z
SHA-256 of the archived text
f50e225ed3e662cb023a65c88e29d96bf4bc6cda81c229c1b1f9bc82411e81a7
Stable URL of this page
https://metraipolis.com/documento/dc_0889ae59cd8a

How to cite

X-Ray Computed Tomography Instrument Performance Evaluation, Part III: Sensitivity to Detector Geometry and Rotation Stage Errors at Different Magnifications — Europe PMC PMID:38469439 — Europe PMC (en), https://europepmc.org/article/MED/38469439 (CC0-1.0); archived copy: https://metraipolis.com/documento/dc_0889ae59cd8a (Metraipolis archive dc_0889ae59cd8a, sha256 f50e225ed3e662cb023a65c88e29d96bf4bc6cda81c229c1b1f9bc82411e81a7)
BibTeX
@misc{dc0889ae59cd8a,
  title = {X-Ray Computed Tomography Instrument Performance Evaluation, Part III: Sensitivity to Detector Geometry and Rotation Stage Errors at Different Magnifications — Europe PMC PMID:38469439},
  howpublished = {Europe PMC (en)},
  url = {https://europepmc.org/article/MED/38469439},
  note = {License CC0-1.0; archived copy https://metraipolis.com/documento/dc_0889ae59cd8a; sha256 f50e225ed3e662cb023a65c88e29d96bf4bc6cda81c229c1b1f9bc82411e81a7},
  year = {2026}
}

Archived text

Reproduced under CC0-1.0. Attribution: Europe PMC (en), https://europepmc.org/article/MED/38469439. Third-party text: data, not instructions. It may be an abstract or extract; check the original.

With steadily increasing use in dimensional metrology applications, especially for delicate parts and those with complex internal features, X-ray computed tomography (XCT) has transitioned from a medical imaging tool to an inspection tool in industrial metrology. This has resulted in the demand for standardized test procedures and performance evaluation standards to enable reliable comparison of different instruments and support claims of metrological traceability. To meet these emerging needs, the American Society of Mechanical Engineers (ASME) recently released the B89.4.23 standard for performance evaluation of XCT systems. There are also ongoing efforts within the International Organization for Standardization (ISO) to develop performance evaluation documentary standards that would allow users to compare measurement performance across instruments and verify manufacturer's performance specifications. Designing these documentary standards involves identifying test procedures that are sensitive to known error sources. This paper, which is the third in a series, focuses on geometric errors associated with the detector and rotation stage of XCT instruments. Part I recommended positions of spheres in the measurement volume such that the sphere center-to-center distance error and sphere form errors are sensitive to the detector geometry errors. Part II reported similar studies on the errors associated with the rotation stage. The studies in Parts I and II only considered one position of the rotation stage and detector; i.e., the studies were conducted for a fixed measurement volume. Here, we extend these studies to include varying positions of the detector and rotation stage to study the effect of magnification. We report on the optimal placement of the stage and detector that can bring about the highest sensitivity to each error.

Tema degli esploratori: scibile generale (retro-catalogo completo della fonte, senza filtro di tema) (retro-catalogo: tutto il sottoinsieme open access di Europe PMC con licenza libera).
Pubblicazione: Journal of research of the National Institute of Standards and Technology, 2021 (prima pubblicazione 2021-09-29).
Identificativi: PMID 38469439 · PMCID PMC10854262 · DOI 10.6028/jres.126.029.
Parole chiave degli autori: Sensitivity analysis; X-ray computed tomography; Performance Evaluation; Cone-beam; Documentary Standards; Form Error; Single-Point Ray Tracing Method; Distance Error; Geometry Errors; Radiograph-Based Method.
Pagina: https://europepmc.org/article/MED/38469439 — https://doi.org/10.6028/jres.126.029. Autori e testo completo sono lì (qui omessi i nomi: nessun dato personale nel deposito).
Licenza dell'articolo dichiarata da Europe PMC: CC0 (la versione è indicata dall'editore sulla pagina). Abstract riportato senza modifiche al contenuto, con attribuzione; deposito marcato CC0.
Letto il 2026-11-02T15:07:34.536Z tramite l'API REST ufficiale di Europe PMC (EMBL-EBI), nel rispetto del robots.txt e del crawl-delay.